  1. Download/Clone PDP from https://github.com/uva-spin/e1039-target-controls (see the PDP folder on this Git repository)
    https://github.com/uva-spin/PDP
  2. Connect all the devices as it was explained in the ORC files on NMR ORC
  3. Install LabView Software
  4. Open "Measurement & Automation Explorer"
  5. Select "Device and Interfaces" > "NI PXle-1071 PXIChassis 1" on the list (see the screenshot below)


  6. Install the required drivers. Notice that there is one "caution" mark appeared in this example for missing driver for "PXI-GPIB" (please see the screenshots below) 


  7.  Rename the PXI names as it appears in the following screenshot


  8.  Select "NMR Tune" mode and check the oscilloscope after tuning the Q-curve


  9. Then you will need to edit the B28_Setup file's "Path" with the correct path where your "events.csv" file will be recorded.


  10.  Click on "Enable Buttons" and click on "New base File" (the one below "Enable Buttons") and then click on the "Disable  Buttons".


  11.  Select "Create a New Baseline"


  12.  Then make sure that you selected "Top Proton" as the target and modify number of sweeps (eg: it will take ~320 seconds for 5000 spweeps; you can reduce/change number on that text box and press "enter" after modification).
     After that, press "One Point" and wait. PDP will take some time (you will see the count-down) and take the measurement (see the bottom left plot)





...