NMR Cables & Coils

to be written.

Lakeshore 218 manual: 218_manual.pdf 

Temperature Sensors

We utilize two types of the temperature sensors;

We plan to arrange a set of sensors per insert as follows;

  • Top cup:       Chip, Chip, Chip, Cernox 
  • Center cup:  None
  • Bottom cup: Chip, Chip, Chip
  • Nose:            Cernox

Readout

  • , Cernox 

Readout

Target Insert: 

The target insert has three PTFE target cups, and each cup is housed with three NMR coils. These NMR coils are started from the top of the target insert, with the SMA connectors routed from inside the shell of the insert and end on the target cups. The top and bottom cups are housed in three-chip resistors, and the Cernox and the middle cup do not have any sensors.  These sensors are reading through the LakeShore 218 via the fisher connector housed at the top of the insert. The fisher (female) connector is attached through the small wires routed inside the insert shell and connected to the sensors on the target cups. To read these sensors, a cable is required that has a fisher (male) connector, and the other end has two DB 25 (male) connectors to connect to the rear panel on the lakeshore DB 25 (female) connectors. For this purpose, the fisher (female) on the insert and the fisher (male) on the cable should be on the same pin configuration. The pin assignment for the target is explained below:

  • These eight These seven sensors are read out by one of LakeShore 218 through one
    • One Fischer connector (

    ??
    • 27 pins) at the insert side and

    two
    • Two D25 (female) connectors at the LakeShore 218 side.

  • The readout cable is of 16 comprises 20 conductors at min minimum (2 wires * 6 sensors chip resistors (12 wires) + 4 wires * 1 sensor).
  • A 24-conductor cable is in place, which goes through the east penetration.
    • The D25 and Fischer connectors are configured for "4 wires * 6 sensors".  Thus they have to be re-configured.
  • A 26-conductor cable will be placed for spare.
  • 2 Cernox (8 wires)). 

Image Added Image Added

Image Added

Image Added 


Mapping between the sensors and the Lakeshore readout: 

Eight sensors (6 chip resistors and 2 Cernox) are connected to the eight channels on the Lakeshore 218 readout. The target cups are numbered from 1 to 9, showing the chip resistors' location as shown in the attached picture. The following table reflects the mapping between the sensors on the target cups and the lakeshore channels. 

Image Added Image Added

Mapping table between the sensors and the Lakeshore channels:

Sensors Cup Lakeshore Channel 
Chip resistor 1Bottom Cup 
Chip resistor 2Bottom Cup2
Chip resistor 3Bottom Cup3
Chip resistor 4Top Cup4
Chip resistor 5Top Cup5
Chip resistor 6Top Cup6
Top Cernox Top Cup7
Bottom CernoxBottom Cup8


View file
nameTarget Insert Cable.pdf
height150
View file
nameTarget Insert Cable.pptx
height150


Spare parts Information

Information relevant to spare parts is mentioned below:

Part/Reference Number Name URLSpare Counts
104497DBEE 105 A102-130 (Female 27 Pole)Product Finder | (fischerconnectors.com)5
104457S 105 A102-130+ (Male 27 Pole)Product Finder | (fischerconnectors.com)5