Comment: Target computer turn-Off/On procedure draft


Turn-Off procedure


The LV code can be just killed in the usual way.


Turn-On procedure


  1. In order to bring it back, one has to find the version of the code which ends with BOS in the e1039 folder (That means it is a latest version where spill ID iteration happens at BOS).
  2. Open the power shell and see if you can establish connection to the EPICS server with regular caget camonitor commands.
  3. Turn on the network disks manually (it is on the right side of the front panel). Please make sure that the NIM bin with the gate generator for BOS/EOS is on. it is in the bottom of the right most rack.
  4. Launch the power shell with camonitor BOS EOS SPILLCOUNTER.
  5. Check the last spill ID generated with caget SPILLCOUNT
  6. Open the LV code. Locate the variable responsible for initial value of the spill ID.
  7. Put in a new number which is bigger than what you got from caget.







This page is dedicated to update the status of the Target Computer Setup at NM4 (kind of a log-book).


Date            Status/Update/Modification                                                                User's name           Additional notes