Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

The fridge is equipped with two valves (run valve and bypass valve) to control the incoming helium from the magnet.  We control the valves using the Cryo Control Panel (CCP) on the target computer.

Fridge Valve Panel in CCP

  • You can open the Panel for the Fridge Valves (Fridge_Valve_Main.vi) by clicking the "Open" button for "Fridge Valve" (Image Modified) on the CCP Main Panel (CCP_Main.vi).
  • Interface for control and monitoring
    • Run valve = Right middle section of the Fridge Valve Panel.
    • Bypass valve = Right bottom section.
  • The number of valve turns is used to specify the opening of valves.
    • Fully closed = 0 turns
    • Fully opened = 3 turns

Image Modified


How to Move Run or Bypass Valve

The operation procedure is common to the run and bypass valves. 

  1. Make sure which valve you handle, and make sure you use the correct interface (i.e. the middle or bottom section).
  2. Decide the set point of valve turns that you move to.
  3. Either
    1. Enter the set point into the "Set Turns" text box () and hit the "Enter" key, or
    2. Press the "Set Turns" button(s) as many as needed ().

In Case of Mistake and Unexpected Situation

  • If you enter a wrong set point, you might press the "Stop Move" button () to stop the valve motion.
  • The set point of valve turns are limited by software from -0.5 to 3.0 turns, where the negative turns are used when expert calibrates the valve position.
  • If you make the set point out of this range somehow, please contact the target expert immediately.