We have two target inserts: Insert # 1 and Insert # 2. Each have insert has two Cernoxs attached to the top and bottom cups, and each Cernox is calibrated. Please check with Target Expert which Insert is loaded in the fridge because each Cernox has a different temperature and resistance curve. Please see below the temperature and resistance curves for each Insert:

Insert # 1: 

1. Insert # 7 (Target Parameter) Cernox Serial Number X130217 installed on the Bottom Cup.

1. Insert # 8 (Target Parameter) Cernox Serial Number X130192 installed on the Bottom Cup. 


Image AddedImage Added










If you see the target parameters, Insert #7 and Insert #8 start lowering down to around 132 Ohms, which is around 50K, then open the Run valve at 0.55 or 0.65 turns to get the liquid in the fridge nose. Keep monitoring the resistance of sensors; it should start increasing after opening the Run valve. Wait for 30 minutes. If the temperature exceeds ~80K (lower than 101 Ohm), open the run valve at two turns, set the separator flow at 30slm, and call the target expert. 

...