Suppose the fridge pressure is above 100 torr. In that case, the Fridge/1000 (Torr) target parameter and the fridge temperature to the Fridge/1000 (K) target parameter are reliable. Suppose the fridge pressure is under 100 torr. In that case, the Fridge/100 (Torr) target parameter is reliable, and the fridge temperature is the Fridge/100 (K) target parameter. 

Image RemovedImage Added


In the above picture, open the 'GV Bypass' VI at point 1, and you can adjust the GV Bypass opening to keep the fridge pressure flat and reduce the LHe consumption. 

...