  1. Confirm that the shim PS is powered ON (if you can go in the hall).
    Image

  2. Press "To Zero", to confirm that CCP can communicate with the magnet PS.


  3. Press the Toggle button of the main switch heater.  It turns on the main switch heater and also the two shim switch heaters.
    1. When you use the PS front panel, you have to press additionally the two buttons of the shim switch heaters on CCP, to turn them on.




  4. Confirm that the two LEDs on the shim PS are ON (if you can go into the hall).
    Image

  5. Set "2 A/m" and "45 A".
    1. You can use the "Presets" selector, or enter the values into the text boxes.


  6. Press "To Set" and wait until the current becomes the set point (SP) and "HOLD".  It takes 23 minutes.


    After ~23 minutes.


  7. Set "1 A/m" and "63 A".


  8. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 18 minutes.


    After ~18 minutes


  9. Set "0.5 A/m" and "74.78 A".


  10. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 24 minutes.


    After ~24 minutes
    Image Added


  11. Switch to the persistent mode, if you want. Follow the steps in the next section.

Reference: UVA elog 154

To Switch to Persistent Mode

  1. Press the Toggle button of the main switch heater.  It turns off both the main and shim switch heaters.
    Image Added

    Image Added

  2. Press "To Zero" and wait until the current becomes 0 A and "HOLD".  It takes 4 minutes, where the ramp rate is slow at the beginning.
    Image Added


To Switch from Persistent Mode

...