Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Confirm that the shim PS is powered ON (if you can go in the hall).
    Image

  2. Press "To Zero", to confirm that CCP can communicate with the magnet PS.


  3. Press the Toggle button of the main switch heater.  It turns on the main switch heater and also the two shim switch heaters.
    1. When you use the PS front panel, you have to press additionally the two buttons of the shim switch heaters on CCP, to turn them on.




  4. Confirm that the two LEDs on the shim PS are ON (if you can go into the hall).
    Image

  5. Set "2 A/m" and "45 A".
    1. You can use the "Presets" selector, or enter the values into the text boxes.


  6. Press "To Set" and wait until the current becomes the set point (SP) and "HOLD".  It takes 23 minutes.


    After ~23 minutes.


  7. Set "1 A/m" and "63 A".


  8. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 18 minutes.


    After ~18 minutes
    Image Added

  9. Set "0.5 A/m" and "74.78 A".
    Image Added

  10. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 24 minutes.
    Image Added

    After ~24 minutes


  11. Switch to the persistent mode, if you want.

...