  1. Confirm that the shim PS is powered ON (if you can go in the hall).
    ImageImage Added

  2. Press "To Zero", to confirm that CCP can communicate with the magnet PS.
    Image Added

  3. Press the Toggle button of the main switch heater.  It turns on the main switch heater and also the two shim switch heaters.
    1. When you use the PS front panel, you have to press additionally the two buttons of the shim switch heaters on CCP, to turn them on.
      Image Added

      Image Added

  4. Confirm that the two LEDs on the shim PS are ON (if you can go into the hall).
    ImageImage Added

  5. Set "2 A/m" and "45 A".
    1. You can use the "Presets" selector, or enter the values into the text boxes.
      Image Added

      Image Added


  6. Press "To Set" and wait until the current becomes the set point (SP) and "HOLD".  It takes 23 minutes.
  7. Set "1 A/m" and "63 A".
  8. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 18 minutes.
  9. Set "0.5 A/m" and "74.78 A".
  10. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 24 minutes.
  11. Switch to the persistent mode, if you want.

...