- Switch back from the persistent mode, if necessary. ** This step is already done in the above section **
- Set "0.5 A/m" and "63 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 24 minutes.
→ after ~24 minutes →
If you are in the NM4 Hall, and you see that the display shows "Rate Limit" in RED, then wait for a few minutes and observe the strip charts and this module's display.
This indicates that the Mercury-iPS is more cautious with the rate of 'current' and it is adjusted by itself.
After a few minutes, the display should look like the following.
Wait until the VI indicates that it switched to "HOLD"
- Set "1 A/m" and "45 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 18 minutes.
→ ~ 18 minutes → - Set "2 A/m" and "0 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 23 minutes.
- Press the Toggle button of the main switch heater. It turns off both the main and shim switch heaters.
...