  1. Switch back from the persistent mode, if necessary. ** This step is already done in the above section **
  2. Set "0.5 A/m" and "63 A".


  3. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 24 minutes.
    → after ~24 minutes →  
    If you are in the NM4 Hall, and you see that the display shows "Rate Limit" in RED, then wait for a few minutes and observe the strip charts and this module's display.
    This indicates that the Mercury-iPS is more cautious with the rate of 'current' and it is adjusted by itself.

    After a few minutes, the display should look like the following.


    Wait until the VI indicates that it switched to "HOLD"

     
  4. Set "1 A/m" and "45 A".
    Image Added
    Image Added

  5. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 18 minutes.
    Image Added → ~ 18 minutes → Image Added

  6. Set "2 A/m" and "0 A".
  7. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 23 minutes.
  8. Press the Toggle button of the main switch heater.  It turns off both the main and shim switch heaters.

...