1) Turning OFF: Push this circular knob (red)
2) To Restart:
1) Rotate the same circular knob like half a turn clock-wise then it will be released (see the picture above)
2) Press the red button on the (right hand) panel to clear the alarm → Press the green button → HR3 will
...
start afterward if there is helium flowing to it.
Reasons for automatic turn-off, and how to restart.
...
Date | Person | Note |
---|---|---|
07/05/2022 | Sam | “ The main reason why the compressor wouldn’t start is if the system is at maximum pressure already or if it did many starts in a short period of time (6 + / hour I think). The system would instantly restart when PC1 > 0.7 even just after it stops due to a lack of pressure rise in PSW4.” |
08/09/2022 | Sam (and Ishara) | Buffer Tank Max Pressure was set to 250 psi (max allowed although Sam asked to change it to 999 psi). |
07/28/2022 | Adam |
|
HR3 LCD screen
Navigation:
F5 > Display Settings
(After pressing F5)
F1 / F2 > Navigate to the Right / Left screens
F3 / F4 > Increase / Decrease the parameter on the selected screen
Press F5 once you are done, to get back to the regular (green) display.
If you want to run the HR3 more or less continuously you could try to do the following.
Increase Sample Time Period and Spacing. (Now it checks pressure increase every 10 minutes, if you set it to an hour, it will only check every hour)
Decrease Max Pressure Rise (Minimum pressure rise detected in buffer tank to keep the HR3 operational, if it's set lower it's more likely that it will keep running 24/7)
...
Ishara changed both the Sample time period (minutes), and Sample Spacing (minutes) to 15 minutes.
Kaeser Compressor
Outside gHe tank pressure:
...