  1. Change the separator flow to 12 SLM (from 7 SLM usually).

  2. Wait for 5 minutes.
  3. Open the bypass valve at 2.0 turns.

  4. Wait for 5 minutes.  Confirm that the fridge return flow goes up.
  5. Change the separator flow to 20 SLM.
  6. Set the bypass valve at 2.5 turns.
  7. Wait until the Ann-Plate Top resistor goes up to 1400 Ohm, which indicates that the liquid started accumulating.Change the separator flow to 20 SLM.

  8. Wait until the Heat-Ex Top resistor goes up to 1400 Ohm, which indicates that the liquid is as high as Heat-Ex Top.

  9. Wait for 1-2 minutes as you need to fill an extra amount of LHe.

  10. Close the bypass valve.

  11. Change the separator flow to 7 SLM.

  12. Make an elog entry to record the time and the mount of this filling.

...