  1. Change the separator flow to 12 SLM (from 7 SLM usually).

  2. Wait for 5 minutes.
  3. Open the bypass valve at 2.0 turns.

  4. Wait for 5 minutes.  Confirm that the fridge return flow goes up.
  5. Set the bypass valve at 2.5 turns.
  6. Wait until the Ann-Plate Top resistor goes up to 1400 Ohm, which indicates that the liquid started accumulating.Set the bypass valve at 2.5 turns.

  7. Change the separator flow to 20 SLM.

  8. Wait until the Heat-Ex Top resistor goes up to 1400 Ohm, which indicates that the liquid is as high as Heat-Ex Top.

  9. Wait for 1-2 minutes as you need to fill an extra amount of LHe.

  10. Close the bypass valve.

  11. Change the separator flow to 7 SLM.

  12. Make an elog entry to record the time and the mount of this filling.

...