Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Set the liquefier to "Manual" mode.

    Image Added
  2. Navigate to the liquefier page on the HMI screen.

  3. Open SVP(A/B).
    Image Added

  4. Navigate back to the "Overview Screen".
    Image Added

  5. Start pre-cool phase:  Open VPC → Open VJV(A/B) → Monitor TX1, TX2, TX3
    Image Added
    1. Charts of TX1, TX2 & TX3:  https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/cool_down.php
    2. TX1 goes down below 100 K in 5 minutes.  If not, there should be a problem.

  6. After ~8mins: 
    1. Open the magnet-bypass valve (i.e. the black hand valve right next to the magnet flow controller),
    2. Open the Magnet Return flow, by changing the flow mode to "OPEN" on the "THCD_401_Main.vi" on the Cryo Control Panel, and
    3. Open VJVT on the HMI screen.

  7. After ~1min: Close VPC and monitor magnet pressure.
    Image Added
    ***
    **************************************************************************************************************************************
    ** Note: If the REGX green hand valve is fixed (check with Target Expert on Shift), then do not change it: just proceed with step #10 *****
    *****************************************************************************************************************************************
    > If you need to change the pressure difference between the magnet and the Dewar then use the REGX all the time to maintain the pressure difference that you need (eg: ~4psi)
  8. After a couple of minutes, you will observe that the magnet pressure drops down suddenly and stay around ~1.4-1.5 psi (this indicates that LHe is flowing to the magnet after the pre-cooling of the last part of the transfer line).
  9. Fill the magnet until the level shows ~44%.
    1. Keep monitoring the liquid amount and the pressure at https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/transfer.php.
  10. Stop the fill if the level reached ~44% or any sudden change occurred during the transfer indicating a problem.
    1. Open VPC
    2. Close VJVT
    3. Close VJV(A/B)
  11. Navigate to the liquefier page on the HMI screen and close SVP(A/B) → Switch the Liquefier to "Liquefy mode".
  12. Wait for 1-3 minutes until the magnet pressure (on the HMI screen) goes down to 1 psi.
  13. Close the Magnet bypass hand valve.
  14. Change the mode of the Magnet Return flow from "OPEN" to "AUTO" on the THCD_401_Main.vi , and then click the "Automation" toggle switch.
  15. Make sure that you leave the VPC open, to capture all the boil-off from the transfer line until it is warm (check TX1). — We need not check TX1 but let VPC open, correct?
  16. Calculate the transfer efficiency of the transfer using https://e906-gat1.fnal.gov/data-summary/e1039/ (see the following screenshot as an example), and write an elog entry on UVA elog.